Meter repairs
April 2022
14. Wait for 30 - 60 seconds to let the coupling fluid spread more evenly on the surface.
15. Insert the transducer capsule into the transducer stalk while aligning the anti-
16. Make sure that the indicator ring on the transducer capsule is flush with the end of
17. Hand-tighten the transducer retainer onto the stalk.
18. Follow
3.7

Transducer cable removal and installation

Rosemount 3410 Series Ultrasonic Gas Flow Meters have red transducer cables that plug
directly in the back of the transformer retainer.
Note
Make a note of the exiting cabling path layout to allow proper tie-wrap configuration later
in this procedure.
Figure 3-15: Rosemount 3410 Series Ultrasonic Gas Flow Meter transducer cables and
ports
A. Cable ties
B. Rosemount 3410 Series Ultrasonic Meter transducer port
C. Transducer assembly
68
Note
Try to avoid pressing the Kapton tape while spreading the coupling fluid.
rotation pin on the capsule with the slot on the inner surface of the stalk.
the transducer stalk to ensure the capsule is set to the correct length (see
3-2).
T-200 transducer housing installation
meter body.
Maintenance and Troubleshooting manual
to install the transducer assembly in a
Emerson.com/Rosemount
00809-0200-3104