Troubleshooting
June 2019
Table 2-1: Troubleshooting
Error
Configuration lost
Electronics Temperature is Out Of Nominal
Range
30
Maintenance and Troubleshooting manual
(continued)
Recommended action(s)
The meter configuration has reset to default
values and the meter is not configured
correctly to measure flow and the meter has
performed a Cold Start.
Unless the Cold Start occurred after
upgrading firmware, replace the CPU
board.
If the cold start occurred after a firmware
upgrade, fully re-configure the meter
from a previously saved configuration
using the Tools → Edit/Compare
Configuration in Daniel MeterLink
Temperature of the electronics is out of
nominal operating range (below -40 °C or
above 100 °C) which could lead to a system
failure.
Attempt to warm or cool the meter
electronics housing.
If the electronics is mounted to the meter
and the process fluid in the meter is over 60
°C, you must remote mount the electronics
off of the meter body.
Collect a Maintenance log using Daniel
MeterLink while the meter is experiencing
the issue and contact your Daniel service
representative.
P/N 3-9000-769
.
Gas Ultrasonic Flow Meters