Troubleshooting
June 2019
Table 2-1: Troubleshooting
Error
Configuration lost
Electronics Temperature is Out Of Nominal
Range
30
Maintenance and Troubleshooting manual
(continued)
Recommended action(s)
•
The meter configuration has reset to default
values and the meter is not configured
correctly to measure flow and the meter has
performed a Cold Start.
—
Unless the Cold Start occurred after
upgrading firmware, replace the CPU
board.
—
If the cold start occurred after a firmware
upgrade, fully re-configure the meter
from a previously saved configuration
using the Tools → Edit/Compare
Configuration in Daniel MeterLink
•
Temperature of the electronics is out of
nominal operating range (below -40 °C or
above 100 °C) which could lead to a system
failure.
•
Attempt to warm or cool the meter
electronics housing.
•
If the electronics is mounted to the meter
and the process fluid in the meter is over 60
°C, you must remote mount the electronics
off of the meter body.
•
Collect a Maintenance log using Daniel
MeterLink while the meter is experiencing
the issue and contact your Daniel service
representative.
P/N 3-9000-769
™
.
Gas Ultrasonic Flow Meters