Emerson Daniel 3410 Series Instructions For Use, Maintenance And Installation Manual page 43

Gas ultrasonic flow meters
Hide thumbs Also See for Daniel 3410 Series:
Table of Contents

Advertisement

Maintenance and Troubleshooting manual
P/N 3-9000-769
Table 2-2: Maintenance
Daniel MeterLink utility
MeterLink Logs/ Reports Menu
Daniel MeterLink Meter
Monitor (Summary) view
Security seals
Maintenance and Troubleshooting manual
(continued)
Diagnostics
Power failure
Chord failure Daniel
MeterLink
Endcap seals
Endcaps latches
Transmitter Electronics
Enclosure
Base Enclosure
Shroud seals
Troubleshooting
June 2019
Action(s)
If this was a known power
fail or restart of the meter
just acknowledge this alarm
on the Status Summary
page.
If this was an unexpected
restart of the meter, verify
the integrity of the power
to the meter and make sure
that the voltage level is in
the range of 11-36 VDC at
the meter.
Collect a complete Archive
log and contact your local
area Daniel service
representative
The meter is unable to
obtain measurement data
from a pair of transducers.
The cause may be isolated
to one pair of transducers
or its cabling. Check the
transducer wiring for this
pair of transducers to make
sure connections are secure
and wired correctly.
Verify the average gain of
this transducer pair is not
above 90dB. Read the value
from the Daniel MeterLink
Monitor Page or using AMS
under Service Tools → Path
performance.
Remove the transducer and
clean the transducer face
( Transducer removal
procedure).
Only authorized personnel
may remove security seals.
Follow your standard
operating procedure to
report seals that have been
tampered with or removed
and replace the seals per
instructions in
External
power source connection
and fuse
in the Installation
Manual (3-9000-765).
43