Troubleshooting
June 2019
Table 2-2: Maintenance
Daniel MeterLink utility
Meter Monitor (Detailed) view
Meter Data List
Meter Monitor (Detailed) view
Meter Data List
40
Maintenance and Troubleshooting manual
(continued)
Diagnostics
•
Electronics Temperature of
out range
•
Electronics voltage out of
range
P/N 3-9000-769
Action(s)
transducer face, or path
length configured
incorrectly).
•
Adjust SSMin or SSMax only
if other checks pass (consult
a Daniel Service
representative before
making these adjustments).
•
Temperature of the
electronics is out of
nominal operating range
below -40 °C or above 100
°C (-40 °F or above 212 °F).
—
Heat or cool the meter
electronics housing. If
operating temperature
exceeds 60oC, remote
mount the Transmitter
Electronics Enclosure.
•
Systems voltages are valid if
1.0V, 1.2V, 2.5V, 3.3V or
the Acquisition Module
valid voltages are 1.2V,
2.5V or 3.3V.
•
Replace the CPU Module if
one or more of the System
Voltages is out of range.
•
Replace the Acquisition
Module if one or more of
the voltages is out of range.
Gas Ultrasonic Flow Meters