Troubleshooting
June 2019
Table 2-2: Maintenance
Daniel MeterLink utility
Meter Monitor (Detailed) view
Meter Data List
Meter Monitor (Detailed) view
Meter Data List
40
Maintenance and Troubleshooting manual
(continued)
Diagnostics
Electronics Temperature of
out range
Electronics voltage out of
range
P/N 3-9000-769
Action(s)
transducer face, or path
length configured
incorrectly).
Adjust SSMin or SSMax only
if other checks pass (consult
a Daniel Service
representative before
making these adjustments).
Temperature of the
electronics is out of
nominal operating range
below -40 °C or above 100
°C (-40 °F or above 212 °F).
Heat or cool the meter
electronics housing. If
operating temperature
exceeds 60oC, remote
mount the Transmitter
Electronics Enclosure.
Systems voltages are valid if
1.0V, 1.2V, 2.5V, 3.3V or
the Acquisition Module
valid voltages are 1.2V,
2.5V or 3.3V.
Replace the CPU Module if
one or more of the System
Voltages is out of range.
Replace the Acquisition
Module if one or more of
the voltages is out of range.