Instructions for use ELECTROmatic TM and TMM/TMC as of panel software version V1.4
6 Operation | 6.3 Regulating the spray water
▶ Press the foot control partly down in order to regulate the speed between
minimum and maximum speed.
6.3 Regulating the spray water
Hazard from insufficient amount of spray water.
Damage to the cog due to overheating.
▶ Make sure that the motor speed is appropriate for the preparation on hand.
▶ Use the requisite minimal amount of spray water.
▶ Rotate the regulating sleeve ② right or left to infinitely adjust the water
portion of the spray.
ð
Regulating sleeve rotated all the way to the right ① = no water supplied
ð
Regulating sleeve rotated all the way to the left ③ = maximum amount of
water
1
3
6.4 PREP mode
Operation general
If the display of the control element shows "PREP", PREP mode is active.
Activated values are shown with a blue background and can be changed.
Deactivated values are shown with a grey background.
▶ Tap the "Lock" control element to unlock settings.
▶ To change a value with a grey background, activate the value by touching
it.
▶ Once the value is activated, use the slider to set the value.
▶ Or press the Plus or Minus key to increase or decrease the activated value
by individual steps.
6.4.1 Changing the speed setting in PREP mode
The user can set 5 speed memories in PREP mode. The speed memories can be
called up by pressing the corresponding control element on the left or right
side. The pre-set speed values are shown on the left or right depending on the
motor. The speeds of the clamped rotating burs are displayed.
CAUTION
2
58 / 90