Hygiene

8 Hygiene

8.1 Tumbler filling and bowl flushing

The following settings can be changed:
▪ Tumbler filling time
▪ Bowl rinsing time
▶ Press the "Tumbler filler" button briefly to start filling the tumbler.
or
▶ Press the "Bowl flush" key briefly to start the bowl flush.
▶ Press the "Set hygiene function" button for more than 4 seconds to start the programming mode.
Set the period of time in 200 ms increments. Minimum: 0.4 s.
An acoustic signal is heard every second.
▶ Release the button to save the selected value.

8.2 Manual hygiene functions of the water-conducting system

Note
RKI recommends rinsing all water-conducting systems at the start of the work day
for 2 minutes (with no handpieces attached) at all dispensing sites, such as, e.g., sy‐
ringe, turbine, scaler motor and tumbler filler.
Note
The "continuous disinfection" hygiene function is available only with the water bottle
kit.
Treatment water that comes into contact with the patient (tumbler, dental instruments,..) can be disin‐
fected manually or intensively disinfected manually.
During normal operation of the treatment centre, the continuous disinfection function (available with
water bottle kit and added disinfectant KaVo OXYGENAL 6) ensures continuous germ reduction of the
water-supplying systems.
Manual rinsing and manual intensive disinfection facilitate the rinsing and disinfection of the water-
supplying system. This counteracts the formation of micro-organisms in times of stagnation and en‐
sures hygienic water supply for the patients.

8.3 Manual intensive sterilisation

Intensive disinfection must be carried out:
▪ At first start-up
▪ Before, during and after prolonged downtimes (more than 4 weeks)
▪ Over the weekend
▪ During vacations
▪ After treatment of patients at high risk
Note
The intensive disinfection bottle must be newly filled prior to every intensive disinfec‐
tion process.
19