Intensive Germ Reduction - KaVo Primus 1058 Life Short Instructions For Use

Hide thumbs Also See for Primus 1058 Life:
Table of Contents

Advertisement

Short instructions for use 10 Hygiene
Set rinsing time
▶ Press the "decrease value" or "increase value" keys to select rinsing program 01
"Programm vorbereiten: 01" or rinsing program 02 "Programm vorbereiten: 02".
ð The standard value for rinsing program 01 is 00:20 minutes per instrument.
ð The standard value for rinsing program 02 is 2:00 minutes per instrument.
▶ Press the foot pedal to display the time for rinsing the instrument hoses and the
M
three-function or multifunctional handpiece.
T
With the foot pedal of the foot control being pressed down, the rinsing time for the re‐
spective rinsing programme can be set in the range from 00:01 to 4:00 minutes using
the "Decrease value" and "Increase value" buttons.
The rinse program proceeds in four steps. For rinse program 01, the cycle starts in step 2, and for
rinse program 02, the cycle starts in step 4.
Start rinsing programme
▶ Briefly press the "Intensive germ reduction" button.
Discontinue the rinsing programme
The rinsing programme can be stopped at any time by pressing the "intensive germ
reduction" key.
If the device is turned off and then on again, the rinsing program also stops.

10.3 Intensive germ reduction

An additional intensive germ reduction must be carried out during extended down‐
times (> 4 weeks).
The need for intensive disinfection is automatically displayed in the top LED of the
"Intensive germ reduction" button at first start-up and after downtimes (weekend, holi‐
days, vacation, etc.).
If an audible signal is issued every 10 seconds and the bottom LED of the "Intensive
germ reduction" button flashes, the KaVo OXYGENAL 6 must be replenished. If a
beep is issued during the filling process (10 times at a one-second interval), the con‐
tainer is full.
Note
If the amalgam separator malfunctions during intensive germ reduction, the entire
process is stopped. The bottom LED of the "HYDROclean" button flashes.
▶ Press the "Intensive germ reduction" button for more than 4 seconds until you hear
an acoustic signal and the upper LED of the "intensive germ reduction" button
flashes. The display switches to the intensive germ reduction menu.