Instructions for use
MASTERmatic LUX M25 L - 1.009.3630
MASTERmatic LUX M05 L - 1.009.3640

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MASTERmatic LUX M25 L and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo MASTERmatic LUX M25 L

  • Page 1 Instructions for use MASTERmatic LUX M25 L - 1.009.3630 MASTERmatic LUX M05 L - 1.009.3640...
  • Page 2 Distribution: Manufacturer: KaVo Dental Technologies, LLC KaVo Dental GmbH Bismarckring 39 11727 Fruehauf Drive Charlotte, NC 28273 USA 88400 Biberach Phone: 847 550 6800 Germany Fax: 847 550 6825 www.kavo.com...
  • Page 3: Table Of Contents

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 Table of contents Table of contents 1 User instructions ....................... 2 Safety..........................2.1 Infection hazard ......................7 2.2 Improper use ........................ 7 2.3 Technical condition ......................7 2.4 Accessories and combination with other equipment............
  • Page 4: User Instructions

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
  • Page 5 Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 1 User instructions Information on the packaging Material number Serial number Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code...
  • Page 6 Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 1 User instructions CAUTION In cases which – if not prevented – can lead to minor or moderate injury. CAUTION In cases which – if not prevented – can lead to material damage.
  • Page 7: Safety

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
  • Page 8: Accessories And Combination With Other Equipment

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 2 Safety | 2.4 Accessories and combination with other equipment 4 Only operate devices or components if they show no signs of damage on the outside.
  • Page 9: Service And Repair

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 2 Safety | 2.6 Service and repair 2.6 Service and repair Repairs, servicing and safety checks may only be performed by trained service personnel. The following persons are authorised to do this: ▪...
  • Page 10: Description Of The Product

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 3 Description of the product | 3.1 Intended use 3 Description of the product The electrical-driven MASTERmatic handpieces are dental handpieces in accord- ance with 21 CFR § 872.4200 (dental handpieces and accessories) designed for use by a trained professional in the field of general dentistry only.
  • Page 11: Technical Specifications

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 3 Description of the product | 3.2 Technical Specifications CAUTION US Federal law restricts this device to sale by or on the order of a health care professional / dentist.
  • Page 12 Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 3 Description of the product | 3.3 Transportation and storage conditions Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi) Protect from moisture (Keep dry)
  • Page 13: Startup And Shut Down

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 4 Startup and shut down | 4.1 Checking the amount of water 4 Startup and shut down WARNING Hazard from contaminated products. Infection hazard to the dentist and patient.
  • Page 14 Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 4 Startup and shut down | 4.1 Checking the amount of water 14 / 28...
  • Page 15: Operation

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING Rotating dental bur. Cuts, infection and burn injury. 4 Never push the press-button while the dental bur is rotating.
  • Page 16: Removing The Medical Device

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 5 Operation | 5.2 Removing the medical device 4 Lightly spray O-rings on the motor coupling with KaVo Spray. 4 Place the medical device on the motor coupling and lock it into place.
  • Page 17: Removing The Dental Bur

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 5 Operation | 5.4 Removing the dental bur CAUTION Dental bur with damaged, worn or deformed shafts. Injury hazard, dental bur may fall out during treatment.
  • Page 18: Checking For Malfunctions And Troubleshooting

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions CAUTION Overheating of the device.
  • Page 19: Changing The Water Filter

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting NOTE The O-rings on the coupling may only be lubricated with a cotton ball wetted with KaVo Spray.
  • Page 20: Reprocessing Steps In Accordance With Iso 17664

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparations at the site of use WARNING Hazard from contaminated products.
  • Page 21: Manual External Disinfection

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing 7.2.3 Manual external disinfection WARNING Incomplete disinfection. Infection hazard. 4 Only use disinfection procedures that are verified to be bactericidal, fungi- cidal and virucidal.
  • Page 22: Automated Reprocessing

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing 7.3 Automated reprocessing WARNING Incomplete disinfection. Infection hazard. 4 Only use disinfection procedures that are verified to be bactericidal, fungi- cidal and virucidal.
  • Page 23: Care Products And Systems - Servicing

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing 7.4 Care products and systems - Servicing WARNING Sharp dental bur in the medical device.
  • Page 24: Packaging

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging NOTE Handpieces must be taken off the service couplings before the chuck service can be started and performed.
  • Page 25: Storage

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 7 Reprocessing steps in accordance with ISO 17664 | 7.7 Storage The medical device has a max. temperature resistance of up to 138 C (280.4 Sterilisation parameters: Sterilisation of the product is approved for the following “wrapped”...
  • Page 26: Optional Aids And Consumables

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. INTRA Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691...
  • Page 27: Terms And Conditions Of Warranty

    Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...