2.3 VOLTAGE SELECTION AND FUSES

The CS Force Tester can be used with electricity supplies of either 230VAC ±10% or
115VAC ±10%. The supply frequency must be 50Hz to 60Hz.
CAUTION: You must check to ensure the correct voltage of your mains
power supply has been selected on the mains input connector BEFORE you
connect mains power to the machine or switch the machine ON.
To check the selected mains voltage, view the red window on the voltage selector. If this is
not set correctly, use a small flat blade screwdriver to open the top of the voltage selector.
The screwdriver should then be used to move the selector from the 115V position to the
230V position as appropriate. You MUST also check that the fuse rating is suitable for the
supply voltage that you are using and the correct fuse values are listed on the serial
number label and also in the table below. The two fuses (live and neutral) are ceramic high
breaking capacity type fuses (20mm x 5mm) and are located inside the voltage selector
assembly.
WARNING: Replace the fuses with the exact same type and rating as the
ones supplied by factory. Do not replace with higher or lower rated fuses.
Fuse Ratings and Part Numbers
Model
Fuse Rating
CS225 - 115VAC
3.15A
CS225 - 230VAC
3.15A
CS1100 - 115VAC
6.3A
CS1100 - 230VAC
6.3A
14 CS User Manual
Part Number
Type
SPK/LS/0011/A
T3A15H250V
SPK/LS/0011/A
T3A15H250V
SPK/LS/0011/B
T6A3H250V
SPK/LS/0011/B
T6A3H250V

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Chatillon cs225