  • Page 1 Reference Manual Compact Signal Calibrator AMETEK JOFRA CSC200  Copyright 2005 AMETEK Denmark A/S...
  • Page 3: Contacting Ametek

    1. Introduction The Ametek Jofra CSC200 is designed to calibrate a wide range of thermocouple and RTD devices. It can source or measure up to 14 different thermocouple (TC) types (including millivolts) and up to 17 RTD types (including ohms). A unique digital "knob" along with selectable decade controls allows quick and easy adjustment of the desired output.
  • Page 4: Safety Information

    1.3 Safety information Symbols Used The following table lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual. Symbol Description AC (Alternating Current) AC-DC Battery CE Complies with European Union Directives Double Insulated Electric Shock Fuse...
  • Page 5 This calibrator must be recycled or disposed of properly (2002/95/EC). The following definitions apply to the terms “Warning” and “Caution”. • “Warning” identifies conditions and actions that may pose hazards to the user. • “Caution” identifies conditions and actions that may damage the instrument being used.
  • Page 6: Calibrator Interface

    Caution To avoid possible damage to calibrator or to equipment under test: • Disconnect the power and discharge all high-voltage capacitors before testing resistance or continuity. • Use the proper jacks, function, and range for your measurement or sourcing application. •...
  • Page 7 Table 1: Key Functions Function Turns the unit on and off Selects between thermocouple or RTD functions Note: Also selects °F/°C units as a power-up function Displays and selects TC or RTD type used in conjunction with knob. Selects input (measure) or output (source) mode Recalls stored setpoints for the function selected Press to store setpoints Left/Right arrow keys move the decade adjust cursor to the desired decade...
  • Page 8: Power Saver

    2.3 Power Saver The calibrator has an adjustable power saver option that allows it to shut down after being dormant for a period of 1 to 30 minutes. To set the shutoff time, perform the following: 1. With the CSC200 turned off, hold down the left arrow key and turn on the power.
  • Page 9 3.1 CJC Control To disable the CJC proceed with the following steps: 1) With the unit turned off press and hold the I/O key. 2) Turn on the unit while still holding down the I/O key. 3) Approximately 5 seconds after power up release the I/O key. The CJC off annunciator will be displayed on the LCD when operated in the TC mode.
  • Page 10 3) Press the "I/O" key to select the input mode. 4) Rotate the knob to select the 2, 3, or 4 wire mode. Be sure the connections made match the input mode selected. Annunciators on the LCD will indicate which wire mode is selected. Figure 3 RTD Output (simulate) Operation To operate the CSC200 in the output mode refer to Figure 4 for proper output...
  • Page 11 Figure 4 5. User Defined Setpoints One of the unique and time saving features of the CSC200 is the ability to store user defined setpoints for all output functions. Up to 3 setpoints can be stored for each of the 14 TC and 17 RTD functions. Stored setpoints will remain in non-volatile memory until changed by the operator.
  • Page 12: Specifications

    4) Setpoint 1 is the starting default location. You can turn the knob to change the location (i.e. Setpoint 1, 2, 3). 5) Once a setpoint location has been chosen, press the "Set" key again to store the output value into that location. 6) Repeat this process for more setpoints (up to three) 7) To recall setpoints simply press the "Recall"...
  • Page 13 RTD Read and Source RTD Type Range (°C / °F) Accuracy (±°C / °F) P10(90)385 -200.0 - 100.0 / -328 - 212 1.4 / 2.52 100.0 - 300.0 / 212 - 572 1.6 / 2.88 300.0 - 600.0 / 572 - 1112 1.8 / 3.24 600.0 - 800.0 / 1112 - 1472 2.0 / 3.60...
  • Page 14 RTD Type Range (°C / °F) Accuracy (±°C / °F) P100(90)392 -200.0 - 100.0 / -328 - 212 0.2 / 0.36 100.0 - 300.0 / 212 - 572 0.3 / 0.54 300.0 - 630.0 / 572 - 1166 0.4 / 0.72 M10(90)427 -100.0 - 260.0 / -148 - 500 1.4 / 2.52...
  • Page 15 Thermocouple Read and Source (errors in °C / °F) TC Type Range (°C / °F) Accuracy (°C / °F) -210.0 - 0.0 / -346 - 32 0.6 / 1.08 0.0 - 800.0 / 32 - 1472 0.4 / 0.72 800.0 - 1200.0 / 1472 - 2192 0.5 / 0.90 -200.0 - 0.0 / -346 - 32 0.8 / 1.44...
  • Page 16: Maintenance

    7. Maintenance 7.1 Replacing Batteries Replace batteries as soon as the battery indicator turns on to avoid false measurements. Note: Use only 9 Volt alkaline battery. 7.2 Cleaning the Unit Warning To avoid personal injury or damage to the calibrator, use only the specified replacement parts and do not allow water into the case.
  • Page 17: Replacement Parts And Accessories

    7.4 Replacement Parts & Accessories Order Number Description SPK-CSC-002 Operating Manual SPK-HHC-002 Soft carrying case 104203 Test Lead Set 124004 Shoulder strap 2206011 Wire adapter kit - Type K thermocouple 2206012 Wire adapter kit - Type T thermocouple 121985 Extension Cable for Pt 100 Sensor - 5 m 121983 Extension Cable for Type K - 5 m 122523...
  • Page 19 SPK-SCS-002 7/05 Rev C 0219130...
  • Page 20 AMETEK is a leading global manufacturer of electrical and electromechanical products for niche markets. Listed on the New York Stock Exchange (AME) since 1930. AMETEK’s annual sales exceed $1 billion. Operations are in North America, Europe, and Asia, with about one third of sales to markets outside the United States.