The HPC50 has the ability to provide a consistent ramp throughout the full mA range.
1
Select
mA
Out.
2
Select the appropriate mA range (0-20, 4-20).
(
)
3
Press the
function
button for Ramp.
4
Enter the 0 and 100% mA points.
5
Enter the total time to get from 0 to 100% of the scale.
(
)
6
function
Start
Press the
button for
and the HPC50 will begin the ramp process.
(
)
7
function
Stop
Press the
button for
to pause the ramp, or
       

MEASURING VOLTAGE

The HPC50 may be used to measure voltages up to 28 VDC. Voltage measurements are limited to the upper display.
To Measure Voltage
X
1
Select
VDC In
.
For more information on working in the HPC50 display, see
2
Connect the HPC50 using the appropriate test leads.
3
The HPC50 will indicate the measured voltage.
Note:
+ OL
The display will indicate
if the measured voltage exceeds 30 VDC.
Back
to go back to the ramp setup screen.