Controlling The Measurement; Chapter 9.5, "Controlling The Measurement - R&S NRPxxS Series User Manual

Three-path power sensors
Table of Contents

Advertisement

®
R&S
NRPxxS(N)
See also
Burst average ("POWer:BURSt:AVG"): The integration time of a measurement is
not predefined but determined by the sensor with the aid of a burst detector. The
start of a burst is detected when the measurement signal rises above a set trigger
level. The measurement ends when the signal drops below a trigger threshold.
See also
Timeslot average ("POWer:TSLot:AVG"): The power is simultaneously measured
in up to 32 time slots. The measurement result is represented by a vector that can
contain up to 32 indices and contains the power of a time slot at each index.
See also
Trace ("XTIMe:POWer"): A sequence of measurements is performed.
See also

9.5 Controlling the Measurement

The power sensor offers a bunch of possibilities to control the measurement:
Do you want to start the measurement immediately after the initiate command or
do you want to wait for a trigger event?
Do you want to start a single measurement cycle or a sequence of measurement
cycles?
Do you want to output each new average value as a measurement result or do you
want to bundle more measured values into one result?
9.5.1 Starting and Ending a Measurement
ABORt
INITiate:ALL
INITiate[:IMMediate]
INITiate:CONTinuous
ABORt
Immediately interrupts the current measurement. If the measurement has been started
as a single measurement (INITiate[:IMMediate]), the sensor goes into the idle
state. However, if a continuous measurement is in progress
ON), the trigger system of the sensor enters the waiting for trigger state, and if the trig-
ger condition is met, a new measurement is immediately started.
See also
Usage:
User Manual 1177.5079.02 ─ 10
Chapter 9.7.1, "Continuous Average
Chapter 9.7.2, "Burst Average
Chapter 9.7.3, "Timeslot Average
Chapter 9.7.4, "Trace
............................................................................................................................ 72
.................................................................................................................... 73
......................................................................................................... 73
........................................................................................................73
Chapter 9.5, "Controlling the
Event
Remote Control Commands
Controlling the Measurement
Measurement",
Measurement",
Measurement",
Measurement",
on page 99.
Measurement",
on page 72.
on page 93.
on page 96.
on page 97.
(INITiate:CONTinuous
72

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents