R&S NRPxxS Series User Manual page 103

Three-path power sensors
Table of Contents

Advertisement

®
R&S
NRPxxS(N)
Measurement result values in the format that is described by the data type (cur-
rently IEEE754 float only)
If no
cuting the measurement, the user data content is finished here. In case that auxiliary
measurands have been selected, the above section is repeated for every auxiliary
measurand. The user data content looks like:
AVGf3100...(400byte AVG values)...MINf3100...(400byte min.
values)...MAXf3100...(400byte max. values)...
Where each of
...(400byte AVG values)...
...(400byte min. values)...
...(400byte max. values)...
Stands for 400 bytes as the equivalent of 100 float values.
Keep in mind that this is only the user data content. It is embedded in the arbitrary
block response data response of the [SENSe<Sensor>:]TRACe:DATA query.
Example:
Usage:
[SENSe<Sensor>:]TRACe:MPWidth?
Queries the attainable time resolution for the Trace mode. The result is the smallest
possible distance between two pixels, i.e. it is the smallest time interval that can be
assigned to a pixel.
Usage:
[SENSe<Sensor>:]TRACe:OFFSet:TIME <time>
Adds an offset to the beginning of the trace sequence. Thus, the trace in the result dis-
play is moved in positive or negative x-direction. If you measure with more than one
sensor, you can use this offset to arrange the traces to each other. The start of record-
ing relative to the trigger event is set using TRIGger:DELay.
Parameters:
<time>
Example:
Manual operation:
[SENSe<Sensor>:]TRACe:POINts <points>
Sets the number of required values per trace sequence.
User Manual 1177.5079.02 ─ 10
[SENSe<Sensor>:]AUXiliary
TRAC:DATA?
Query only
Query only
Range:
*RST:
Default unit: Seconds
TRAC:OFFS:TIME 1.0
See
"Trace Offset Time"
Remote Control Commands
Configuring the Measurement Modes
measurands have been activated before exe-
- (500 ms + TRGger:DELay) to 100 s
0.0
on page 48