System:serror:list[:Next]; System:sutime; System:tlevels; System:transaction:begin - R&S NRPxxS Series User Manual

Three-path power sensors
Table of Contents

Advertisement

®
R&S
NRPxxS(N)

SYSTem:SERRor:LIST[:NEXT]?

Queries the list of all static errors that have occurred but have already been resolved
for the oldest entry and removes it from the queue. The response consists of an error
number and a short description of the error.
Example:
Usage:
SYSTem:SUTime <update_time>
Effective only in the NRP legacy mode.
Sets the status update time. That is the maximum rate in which the power sensor can
output measurement results. Relevant only in continuous measurement mode
(INITiate:CONTinuousON).
Parameters:
<update_time>

SYSTem:TLEVels?

Queries the possible power test levels of the sensor.
Usage:

SYSTem:TRANsaction:BEGin

Starts a series of settings.
Usage:
SYSTem:TRANsaction:END
Ends a series of settings.
Usage:
SYSTem[:SENSor]:NAME <sensorname>
Sets the name of the sensor according to your requirements. The specified name is
displayed in the web user interface of the network sensors.
User Manual 1177.5079.02 ─ 10
SYST:SERR:LIST?
Query
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification"
Response
Query only
Range:
0.0 to 10.0
*RST:
10e-3
Default unit: Seconds
Query only
Event
Event
Remote Control Commands
Configuring the System
139

Hide quick links:

Advertisement

Table of Contents
loading