R&S NRPxxS Series User Manual page 99

Three-path power sensors
Table of Contents

Advertisement

®
R&S
NRPxxS(N)
Manual operation:
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh <width>
Sets the length of the timeslot (see
Parameters:
<width>
Manual operation:
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:OFFSet[:TIME] <time>
Determines the distance from the start of the timeslots to the start of the interval to be
blanked out (see
Parameters:
<time>
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:TIME <time>
Sets the length of the time interval in the timeslots to be excluded from the measure-
ment (see
Note: Even if the exclusion interval exceeds the timeslot because, for example, its right
limit is outside the timeslot, correct results are obtained. In the extreme case, where
the interval length has been set to a value greater than the timeslot length, 0 W is out-
put as the measured power. No error message is output.
Parameters:
<time>
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID[:STATe] <state>
Enables or disables the blanking out of time intervals in the timeslots.
Parameters:
<state>
9.7.4 Trace Measurement
The trace measurement is used to acquire the course of power over a certain time.
During the measurement time ([SENSe<Sensor>:]TRACe:TIME) a large number of
measurements are made and the result is returned to the user as an array of values
User Manual 1177.5079.02 ─ 10
See
"Number of Timeslots"
Range:
*RST:
Default unit: Seconds
See
"Nominal Width"
Figure
9-3).
Range:
*RST:
Default unit: Seconds
Figure
9-3). The parameter applies to each individual timeslot.
Range:
*RST:
Default unit: Seconds
*RST:
Remote Control Commands
Configuring the Measurement Modes
on page 46
Figure
9-3).
10.0e-6 to 0.10
1.000e-3
on page 47
0.00 to 0.10
0.00
0.00 to 0.10
0.00
OFF