Trigger:master:state; Trigger:slope; Trigger:source; Trigger:sync:port - R&S NRPxxS Series User Manual

Three-path power sensors
Table of Contents

Advertisement

®
R&S
NRPxxS(N)
TRIGger:MASTer:STATe <state>
Enables or disables the trigger master mode of the sensor. In this state, the power sen-
sor can output a digital trigger signal in sync with its own trigger event.
If enabled, select the output port for the trigger signal using TRIGger:MASTer:PORT.
Typically, the trigger master uses its internal trigger source. But you can also trigger the
trigger master externally, because the power sensor has got two external trigger con-
nectors. If you trigger the trigger master externally, use EXTernal1 as external trigger
input port (trigger source) and EXTernal2 as trigger master output port or vice versa.
Parameters:
<state>
TRIGger:SLOPe <slope>
Effective only if
Determines which edge of the envelope power, with internal triggering, or increasing
voltage, with external triggering, is used for triggering.
Parameters:
<slope>
Manual operation:
TRIGger:SOURce <source>
Selects the source for the trigger event detector.
Parameters:
<source>
Manual operation:
TRIGger:SYNC:PORT <sync_port>
Selects the external connection for the sync output of the sensor. For more information,
see TRIGger:SYNC:STATe.
User Manual 1177.5079.02 ─ 10
*RST:
OFF
is set to INTernal or EXTernal.

TRIGger:SOURce

POSitive | NEGative
POSitive
Rising edge
NEGative
Falling edge
*RST:
POSitive
See
"<Slope>"
on page 54
HOLD | IMMediate | INTernal | BUS | EXTernal | EXT1 |
EXTernal1 | EXT2 | EXTernal2
See
Chapter 9.5.2.3, "Trigger
*RST:
IMMediate
See
"<Source>"
Remote Control Commands
Controlling the Measurement
Sources",
on page 54
on page 74.