Getting Started

2.5 Connecting Electrodes and Temperature Sensor

Align the BNC connector slots with the posts
2.5mm Phono plug for
temperature measurement
of meter's socket and rotate connector
clockwise until it locks - do not force. To
BNC
connector of
remove, rotate the connector in counter-
pH, ORP, or
clockwise direction until it unlocks, and slide
Ion Selective
Electrode
the connector off the socket. Insert the mini
phono jack of temperature sensor into the
socket on the meter. Unplug the phono
jack to perform measurements without
temperature compensation (25°C default).

2.6 Powering the Meter On

Press ON/OFF. All LCD segments will display momentarily during the self-
diagnostic test, before scrolling the model number, software revision, pH
buffer group, and number of pH calibration points selected before returning
to the measurement mode:
If a temperature probe is not connected, either 25.0ºC (factory default) or the
last calibrated temperature value is used. If a temperature probe is connected,
the current measured temperature is used.
"Or" (Over range) and "Ur" (Under range) indicates the reading exceeds the
maximum or minimum measurement range. See Section 6 - Specifications.
However, the most likely reason for these error messages is that the electrode
is not connected or broken.