Waters ACQUITY UPLC H-Class Series System Manual page 66

Table of Contents

Advertisement

4.
On the Equilibrate to Method tab, click each module sub-tab and if necessary, change
the settings for the flow rate, solvent composition, temperature, and lamp state to match
your requirements at equilibration.
Table 3–14: Equilibrate to Method table values
System startup
parameters
Method initial flow rate 0.500 mL/min
Composition of A, B,
C, and D (sum must
be 100%)
Column temperature
Sample temperature
Lamp
Note:
Change the settings on the Optional Characterize tab only if the needle has been
replaced.
5.
Click Start.
Result:
1.
The system begins to prime.
2.
The system establishes the method flow rate, column and sample temperatures, and
ignites the lamp.
3.
After priming, the sample manager characterizes the needle and seal, if selected,
and then logs the results of the characterizations into the database.
Default
Allowed values
0.1 to 2.0 mL/min
A, 100%
A; 0 to 100%
B, C, D, 0%
B; 0 to 100%
C; 0 to 100%
D; 0 to 100%
Off
Depends on type of column compartment
Note:
Column
selection for the
Column Manager
defaults to Column
1.
On
Off, or 4.0 to 40.0 °C
On
On or Off
Note:
power-on, operate, or ignite the lamp of the
detector when there is no flow through the
cell, or when the cell is dry.
April 4, 2018, 715005702 Rev. A
Page 66
For light guiding flow cells, do not

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents