Figure 2–8: Removing ferrule from fitting
#1 knurled compression fitting
Ferrule (captive ferrule shown)
4.
Discard the used ferrule.
5.
Install the new ferrule on the fitting.
6.
Slide the #1 knurled compression fitting and ferrule onto the tubing.
Figure 2–9: Sliding fitting and ferrule onto tubing
Tubing
#1 knurled compression fitting
Note:
If it is a captive ferrule, the ferrule locks into the fitting.
7.
Screw the #2 locking cap nut onto the #1 knurled compression fitting.