Entering Run Time; Entering A Timed Run - Beckman Coulter Avanti J-26S XPI Instructions For Use Manual

High-performance centrifuge
Table of Contents

Advertisement

The
RPM = - -  G
ROTOR ID
3
Enter the required RCF on the keypad and press
• The RCF and the corresponding RPM appear in the Speed: field.
• The RCF ( g) indicator illuminates on the actual value display, indicating that RCF ( g) is
being displayed during the run.
When
• If you enter an RCF value that is unachievable for the installed rotor, a message will appear
telling you the rotor's achievable RCF.

Entering Run Time

Run time can be entered in hours and minutes or in 
is used for continuous operation.

Entering a Timed Run

1
Press
TIME
HH:MM:
2
Use they keypad to enter the required time (up to 99 hours, 59 minutes) and press
PN B10093AE
field flashes.
SPEED
RPM/RCF
RCF (x g)
Rotor : JA-30.50
Speed : - -
RPM
= 81 800 x G
Time : 00:00
Temp : 25°C
Max Temp : 29°C
Accel : MAX
Decel :
and
are pressed, the centrifuge runs at the calculated RPM.
ENTER
START
.
,
and
soft keys appear.
W2T,
HOLD
TIME
TEMP
HR:MIN
RPM
RCF
.
ENTER
2
t (accumulated centrifugal effect). Hold mode
Operation
Manual Operation
°C
.
ENTER
2
2-7