Configuration and Use Manual
MMI-20048166
Table 7-2: Options for alerts and conditions
Alert
Default severity: Function Check
Severity configurable: No
Fault Timeout applicable: No
Sensor Being Simulated
Default severity: Function Check
Severity configurable: No
Fault Timeout applicable: No
Output Fixed
Default severity: Function Check
Severity configurable: No
Fault Timeout applicable: No
Drive Over-Range
Default severity: Maintenance
Required
Severity configurable: Yes
Fault Timeout applicable: Yes
Process Aberration
Default severity: Out of
Specification
Severity configurable: Yes
Fault Timeout applicable: Yes
Configuration and Use Manual
(continued)
Name
[131] Meter Verification in
Progress
[132] Sensor Simulation
Active
[101] mA Output 1 Fixed
[114] mA Output 2 Fixed
[111] Frequency Output 1
Fixed
[118] Discrete Output 1 Fixed
[102] Drive Overrange
[105] Two-Phase Flow
[115] External Input Error
[121] Extrapolation Alert
(Concentration)
[116] Temperature Overrange
(API referral)
[117] Density Overrange (API
referral)
[123] Pressure Overrange (API
referral)
Configure device options and preferences
Conditions
Description
A meter verification test is running.
Sensor simulation mode is enabled.
(Enhanced core processor only)
The HART address is set to a non-
zero value, a loop test is running, or
the output is configured to send a
constant value (mA Output Action
or Loop Current Mode).
The output is configured to send a
constant value. A loop test may be in
progress.
The output is configured to send a
constant value. A loop test may be in
progress.
The output is set to a constant state.
A loop test may be in progress.
The drive power (current/voltage) is
at its maximum.
The line density is outside the user-
defined two-phase flow limits.
The connection to an external
measurement device has failed. No
external data is available.
The line density or line temperature
is outside the range of the
concentration matrix plus the
configured extrapolation limit.
The line temperature is outside the
range of the API table.
The line density is outside the range
of the API table.
The line pressure is outside the
range of the API table.
April 2019
Ignorable
Yes
No
Yes
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
91