Emerson Micro Motion 4200 Configuration And Use Manual page 88

Hide thumbs Also See for Micro Motion 4200:
Table of Contents

Advertisement

Configure device options and preferences
April 2019
Important
The transmitter reports all the process and device conditions that were reported by previous transmitters.
However, the transmitter does not report them as individual alerts. Instead, the transmitter reports them as
conditions associated with alerts.
Procedure
To change the severity of an alert:
a) Choose Configure > Alert Setup.
b) Choose the category of the alert: Sensor, Configuration, Process, or Output.
c) Select the alert.
d) Set the severity as desired.
Option
Failure
Function Check
Out of Specification
Maintenance Required
To ignore an alert:
a) Choose Configure > Alert Setup.
b) Choose the category of the alert: Sensor, Configuration, Process, or Output.
c) Select the alert.
d) Set the severity to No Effect.
If an alert is ignored, any occurrence of this alert is not posted to the alert list and the status LED on the
transmitter does not change color. The occurrence is posted to alert history.
To ignore a condition:
a) Choose Configure > Alert Setup.
b) Choose the category of the alert: Sensor, Configuration, Process, or Output.
c) Select the alert.
d) Choose Set Conditions.
e) Select the condition and set it to OFF.
If a condition is ignored, any occurrence of this condition is not posted to the alert list and the status LED
on the transmitter does not change color. The occurrence is posted to alert history.
88
Description
The event is serious enough to require fault actions by the transmitter. The event
may be either device-related or process-related. Operator action is strongly
recommended.
Configuration change or device testing. No fault actions are performed. The
operator may need to complete a procedure.
The process is outside user-specified limits or device limits. No fault actions are
performed. The operator should check the process.
Device maintenance is recommended, either near-term or mid-term.
Configuration and Use Manual
MMI-20048166
Micro Motion 4200 Transmitters