April 2019
Table 12-1: SMV abort codes
Code
Description
10
State in progress
11
Verification Complete
12
Transmitter in fault and SMV
cannot run
13
No factory air verification
14
No field reference point
15
Drive frequency drift from carrier
frequency
Other
General abort
12.1.3 Set up SMV automatic execution
You can set up and run a single test at a user-defined future time. You can also set up and run tests
automatically on a regular schedule.
Automatic execution of SMV is managed from the transmitter. You do not need a connection from an
external configuration tool.
Tip
The time between test runs must be between 1 and 1000 hours. The time to the first test run can be any
positive floating number.
Important
Only the 20 most recent SMV results are stored. For SMV Basic versions, the last automatic execution result
can be viewed, while all 20 results can be viewed with SMV licensed. To view or chart these results using an
external tool, you must upload them from the transmitter.
Related information
Set up SMV automatic execution using the display
Set up SMV automatic execution using ProLink III
Set up SMV automatic execution using the Field Communicator
Set up SMV automatic execution using the display
Procedure
1. Choose Menu > Operations > Smart Meter Verification > Schedule Verification.
2. To schedule a single test:
a) Set Hours to 1st Run to the number of hours to elapse before the test is run.
b) Set Hours Between to 0 .
3. To schedule a recurring execution:
a) Set Hours to 1st Run to the number of hours to elapse before the first test is run.
130
(continued)
Recommended actions
NOT AVAILABLE
NOT AVAILABLE
View the Alerts present on the device and take any necessary
actions to clear them before proceeding.
Perform factory calibration on air.
Perform a field reference point procedure.
Ensure temperature, flow, and density are stable. Start again.
Repeat the test. If the test aborts again, contact customer
service.