Configure device options and preferences
April 2019
Table 7-2: Options for alerts and conditions
Alert
Severity configurable: No
Fault Timeout applicable: Yes
Configuration Error
Default severity: Failure
Severity configurable: No
Fault Timeout applicable: Yes
Tube Not Full
Default severity: Failure
Severity configurable: No
Fault Timeout applicable: Yes
Extreme Primary Purpose Variable
Default severity: Failure
Severity configurable: No
Fault Timeout applicable: Yes
Transmitter Initializing
Default severity: Failure
Severity configurable: No
Fault Timeout applicable: No
Function Check in Progress
90
(continued)
Name
[016] Sensor Temperature
(RTD) Failure
[017] Sensor Case
Temperature (RTD) Failure
[020] Calibration Factors
Missing
[021] Incorrect Sensor Type
[030] Incorrect Board Type
Password Not Set
Time Not Entered
[120] Curve Fit Failure
(Concentration)
[033] Insufficient Pickoff
Signal
[005] Mass Flow Rate
Overrange
[008] Density Overrange
[009] Transmitter Initializing/
Warming Up
[104] Calibration in Progress
Configuration and Use Manual
Conditions
Description
The value computed for the
resistance of the line RTD is outside
limits.
The values computed for the
resistance of the meter and case
RTDs are outside limits.
Some calibration factors have not
been entered or are incorrect.
Transmitter verification of sensor
circuits and characterization has
produced a discrepancy. The
transmitter cannot operate the
sensor.
The firmware or configuration
loaded in the transmitter is
incompatible with the board type.
Display security has been enabled
but the display password has not
been changed from the default
value.
The system time has not been
entered. The system time is required
for diagnostic logs.
The transmitter was unable to
calculate a valid concentration
matrix from the current data.
The signal from the sensor pickoffs is
insufficient for operation. (Enhanced
core processor only)
The measured flow is outside the
sensor's flow limits.
The measured density is above
10 g/cm³.
The transmitter is in power-up
mode.
A calibration is running.
Micro Motion 4200 Transmitters
MMI-20048166
Ignorable
No
No
No
Yes
No
No
Yes
No
Yes
No
No
No
No