If your health care professional has enabled Ramp on
your ventilator, you can adjust the Ramp Start Pressure
setting. The setting can be adjusted from 4.0 to your
prescription pressure setting in increments of 1.
Flex – The optional Flex feature allows you to adjust the
level of air pressure you feel when you exhale during
therapy. If this feature is enabled by your health care
professional, you can adjust the setting from 1 to 3.
Trilogy100 clinical manual
Note: If the Flex feature is
enabled, the Rise Time setting
will not appear on your My
Settings screen and cannot be
adjusted. When Flex is enabled,
the Rise Time is fixed at a setting
of 3.
Note: When adjusting the Flex
setting, it is recommended that
you start with the minimum
setting of 1, which provides
the least relief. Levels 2 and
3 progressively increase the
pressure relief.
Note: If you are using an
active circuit, the Flex feature is
unavailable.
Note: Flex is only available if
Auto-Trak is enabled.