76
AVAPS
AVAPS is only available if Flex is not enabled.
You can disable AVAPS by selecting Off, or you can enable AVAPS by
selecting On. If you select Off, the IPAP setting displays. If you select
On, the IPAP Max Pressure and IPAP Min Pressure display.
IPAP
The IPAP setting displays if AVAPS is Off. You can increase or decrease
the Inspiratory Positive Airway Pressure (IPAP) from 4 to 50 cm H
in increments of 1. IPAP is limited to a maximum of 25 cm H
Flex is enabled. You cannot set the IPAP setting lower than the EPAP
setting.
IPAP Max Pressure
The IPAP Max Pressure setting displays if AVAPS is enabled. You can
increase or decrease the setting from 4 to 50 cm H
of 1. The IPAP Max Pressure must be equal to or greater than the IPAP
Min value.
IPAP Min Pressure
The IPAP Min Pressure setting displays if AVAPS is enabled. You can
increase or decrease the setting from 4 to 50 cm H
of 1. The IPAP Min Pressure must be equal to or greater than the EPAP
value, and it must be less than or equal to the IPAP Max Pressure.
EPAP
You can increase or decrease the Expiratory Positive Airway Pressure
(EPAP) from 4 to 25 cm H
EPAP can be set to zero.
When AVAPS is disabled, the EPAP setting must be less than or equal
to the IPAP setting. When AVAPS is enabled, the EPAP pressure must
be less than or equal to the IPAP Min Pressure.
Trilogy100 clinical manual
O in increments of 1. For active circuits,
2
O
2
O when
2
O in increments
2
O in increments
2
Note: AVAPS is only available if
the Passive Circuit is selected.
Note: IPAP, IPAP Max, or IPAP
Min cannot be set to more than
30 cm H
O above EPAP.
2
Note: EPAP cannot be set to
more than 30 cm H
O below
2
IPAP, IPAP Max, or IPAP Min.