3. Verify that the rubber feet are on the bottom of the device.

Initial Setup

1. Connect the power cord to the device and then to an AC outlet.
2. Attach the test lung to the patient connection end of the desired
circuit (Active PAP or Passive).
3. Follow the instructions in Chapter 5 to access the Setup Screen.

Settings and Alarms Tests

Complete the following steps to set up the settings and alarms tests.
1. Setup
A. Settings And Alarms Menu
Modify the settings in the Settings and Alarms menu to match
those shown below in Table 1. If necessary, refer to Chapter 5 for
instructions on modifying ventilator settings.
Setting
Dual Prescription
Circuit Type
Therapy Mode
AVAPS (passive circuit only)
IPAP
EPAP
Breath Rate
Inspiratory Time
Trigger Type (passive circuit)
Flow Trigger Sensitivity (active
PAP circuit)
Flow Cycle Sensitivity (active
PAP circuit)
Trilogy100 clinical manual
Value
Off
Active PAP or Passive
S/T
Off
20 cm H
O
2
4 cm H
O
2
12 BPM
1.6 seconds
Auto-Trak
6.0 l/min
20%
Table 1 - Ventilator Settings
in the Settings and Alarms
Menu