Measured Patient Parameters
Parameter
Vte/Vti
Minute Ventilation
Estimated Leak Rate
Respiratory Rate
Peak Inspiratory Flow
Peak Inspiratory Pressure
Mean Airway Pressure
% Patient Triggered Breaths
I:E Ratio
All flows and volumes are measured at BTPS conditions.
Spontaneous Breathing During Power Failure Conditions
Flow Set Point
(l/min)
30
60
Disposal
Dispose of the device and batteries in accordance with local regulations.
Trilogy100 clinical manual
Range
0 to 2000 ml
0 to 99 l/min
0 to 200 l/min
0 to 80 BPM
0 to 200 l/min
0 to 99 cm H
0 to 99 cm H
0 to 100%
9.9- 1: 1-9.9
Inspiratory Resistance (cm H
Active Circuit
Passive Circuit
< 2.0
< 10.0
Accuracy
Greater of 15ml or 15% of reading
Calculation based on measured Vte or Vti
and Respiratory Rate
N/A
Greater of 1 BPM or 10% of reading
3 l/min plus 15% of reading
O
Greater of 2 cm H
2
O
Greater of 2 cm H
2
N/A
Calculation based on Inspiratory Time and
Expiratory Time
O)
Expiratory Resistance (cm H
2
Active Circuit
< 1.0
< 1.5
< 4.0
< 4.0
O or 10% of reading
2
O or 10% of reading
2
O)
2
Passive Circuit
< 1.2
< 3.7