Alarm
Priority
Battery Not
Info
Charging
Check External
Info
Battery
Battery
Info
Depleted
Audible
Visual (Alarm
Indicator Button
and Display)
"Detach Battery Not
Charging" or "Internal
Battery Not Charging"
message
"Check External
Battery" message
"External Battery
Depleted" or
"Detachable Battery
Depleted" message
appears
Device
User/Clinician Action
Action
User: Press Alarm Indicator/
Operates
Audio Pause button to
silence alarm. Or, press
Left (Reset) button to reset
alarm. If condition continues
for internal battery, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: If condition
continues for internal battery,
place patient on alternate
source of ventilation and
have device serviced.
User: Press Alarm Indicator/
Operates
Audio Pause button to
silence alarm. Or, press
Left (Reset) button to reset
alarm. Check connection to
external battery. Replace
external battery with another
external battery, if available.
If alarm continues, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: If the alarm
continues after the measures
above have been taken, place
patient on alternate source of
ventilation and have device
serviced.
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Replace depleted battery
with another or switch to AC
power, if available.
Clinician: Replace depleted
battery with another or
switch to AC power.