PHILIPS Respironics Trilogy 100 Clinical Manual page 33

Table of Contents

Advertisement

26
During the time window, if patient effort is detected, either a
Spontaneous or an Assist breath is delivered for the first effort. If the
last breath delivered was Mandatory, then a Spontaneous breath is
delivered for the first patient effort of the window. If the last breath
delivered was an Assist or Spontaneous breath, then an Assist breath
is delivered for the first patient effort of the window. Assist breaths
deliver the Inspiratory Pressure setting during inhalation and the
PEEP setting during exhalation. Pressure supported spontaneous
breaths are delivered above the PEEP setting (PEEP compensated).
During the time window, if patient effort continues to be detected
(after the first patient triggered breath), Spontaneous breaths are
delivered for the remainder of the window. The figure below provides
example breath patterns in PC-SIMV mode.
Inspiratory Pressure = 26 cm H
Pressure Support = 10 cm H
PEEP = 6 cm H
O
2
Rate = 5 BPM
12 seconds
PRESSURE
M
S
26
20
10
2
1
1
6
Enter PC-SIMV mode
5
10
Trilogy100 clinical manual
O
= Pressure Supported
1
2
O
Breath
2
= Mandatory Breath
2
= Assist Breath
3
12 seconds
12 seconds
M
S
M
3
1
15
20
25
30
= Mandatory Window
M
= Spontaneous Window
S
12 seconds
M
2
35
40
45
TIME
Sample Breath Patterns in
PC-SIMV Mode