Battery Function Verification

Make sure the batteries are functioning properly and fully charged
before patient use.
1. Verify the Detachable and Internal (Lithium-Ion)
Batteries Function
A. Connect AC Power to the device and verify that the green AC
LED on the front panel is lit.
B. Verify that the detachable battery is properly installed.
C. Turn the device on and verify that both the detachable and
internal battery symbols appear on the display. Verify that if
either battery is less than fully charged, the charge symbol will
display on the respective battery.
D. Disconnect the AC Power source from the device.
Verify that the AC Power Disconnected alarm
message appears on the display and the green AC
LED is not lit. Press Reset.
Verify that the detachable battery symbol shows the
level of charge noted in the previous step and that
the device continues to operate.
Verify that the detachable battery symbol has a black
box around it to indicate that it is in use.
E. Disconnect the detachable battery pack from the device.
Verify that the Detach Batt Disconnected alarm
message appears on the display. Press Reset.
Verify that the internal battery symbol shows the
same level of charge as noted in Step C and the
device continues to operate.
Verify that the internal battery symbol has a black
box around it to indicate that it is in use.
F.
Reconnect the Detachable Battery and AC Power source.
Trilogy100 clinical manual