Cleaning The Reusable Exhalation Devices; Replacing The Bacteria Filter (Optional) - PHILIPS Respironics Trilogy 100 Clinical Manual

Table of Contents

Advertisement

5. Place the circuit on a clean towel to dry. Do not wipe dry. The
circuit must be completely dry before storing.
6. Reassemble the circuit when dry. Store in a plastic bag or dust-
free area.
7. Inspect components for deterioration prior to use.

Cleaning the Reusable Exhalation Devices

Disassemble the exhalation device from the patient circuit. Follow
the detailed cleaning instructions included with your exhalation
device.

Replacing the Bacteria Filter (Optional)

To remove a bacteria filter and replace it with a new one:
1. Disconnect the flexible tubing from the bacteria filter by gently
pulling straight out from the filter. The proximal tubing (if using
the Active PAP Exhalation Device) can remain connected to the
ventilator.
2. Remove the used bacteria filter from the ventilator by
gently pulling straight out from the connector. Follow the
manufacturer's labeling and your institution's infection control
guidelines for disposal of the bacteria filter.
3. Attach a new bacteria filter. The openings on either end of the
bacteria filter are different sizes. Fit the larger opening onto the
gas outlet port on the side of the ventilator as shown.
WARNING
Philips Respironics recommends
that a main line outlet bacteria
filter (Part Number 342077) be
used whenever the device is
used for invasive therapy or if
the ventilator may be used on
multiple patients.
Chapter 7 Cleaning and Maintenance