Changing The Exhalation Porting Blocks - PHILIPS Respironics Trilogy 100 Clinical Manual

Table of Contents

Advertisement

Exhalation Valve Port
Proximal
Pressure
Port
Exhalation Valve Line
Refer to the instructions included with the Active Exhalation
Device for more information.

Changing the Exhalation Porting Blocks

The Passive Exhalation Porting Block is pre-installed on the device. If
you need to change the exhalation porting blocks, follow the steps
below.
1. Using a screwdriver, remove the screw located in the center of
the exhalation porting block, as shown below.
Exhalation Porting Block
Proximal
Pressure
Line
Exhalation Valve Port
on Porting Block
Connecting an Active
Exhalation Device with
Proximal Pressure
Unscrewing the Exhalation
Porting Block
Chapter 4 Ventilator Setup
55