Therapy Mode Table - PHILIPS Respironics Trilogy 100 Clinical Manual

Table of Contents

Advertisement

Therapy Mode Table

The following table summarizes all of the therapy modes and the settings available in each mode.
Some settings in the table are dependent upon other settings. For instance, if Trigger Type is set to
Flow Trigger, then the Flow Trigger Sensitivity and Flow Cycle settings will display.
Note: Pressure Support, referred to in the table below and later in this manual, is defined as IPAP - EPAP or Pressure -
PEEP (PC-SIMV).
Dual Prescription
Circuit Type
CPAP
IPAP
AVAPS (On, Off )*
IPAP Max Pressure
IPAP Min Pressure
EPAP
Inspiratory Pressure
Pressure Support (PS)
PEEP
Tidal Volume
Breath Rate
Inspiratory Time
Trigger Type*
Flow Trigger Sensitivity
Flow Cycle Sensitivity
Ramp Length
Ramp Start Pressure
Flex *
Rise Time
Flow Pattern
Sigh
Circuit Disconnect
Apnea
Apnea Rate
High Vte*
Low Vte*
High Vti*
Low Vti*
High Minute Ventilation
Low Minute Ventilation
High Respiratory Rate
Low Respiratory Rate
High Inspiratory Pressure
Low Inspiratory Pressure
* AVAPS, Trigger Type, Flex, High Vte and Low Vte settings are only available with the Passive circuit type. Trigger Type is not
available with the Passive circuit in CPAP mode. High Vti and Low Vti settings are only available with the Active circuit with PAP.
CPAP
S
S/T
Therapy Modes
T
PC
PC-SIMV
Chapter 3 Modes, Features, and Alarms