102
Alarm
Priority
Ventilator
High
Service
Required
Check Circuit
High
Trilogy100 clinical manual
Audible
Visual (Alarm
Indicator Button
and Display)
• • • • •
Red flashing button;
"Ventilator Service
Required" message
• • • • •
Red flashing button;
"Check Circuit"
message
Device
User/Clinician Action
Action
User: Press Alarm Indicator/
Operates
Audio Pause button to silence
alarm. Or, press Left (Reset)
button to reset alarm. The
alarm goes away until next
reminder appears. Connect
patient to alternate source of
ventilation and contact your
health care professional as
soon as possible.
Clinician: If alarm continues,
have the device serviced.
User: Press Alarm Indicator/
Operates
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Make certain the circuit type
matches what is selected in
the Setup Menu. Make sure
there is no kinked or pinched
tubing. Make sure tubing is
properly attached. Make sure
all tubing and exhalation
devices are clean and free
of condensation (refer to
the Troubleshooting section
– tubing condensation).
If using Active Exhalation
Device, make sure diaphragm
is inserted correctly and is
not wrinkled or kinked. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status. Verify that breathing
circuit setup is correct.
Correct any errors. If alarm
continues, have device
serviced.