Alarm Summary Table

The following table summarizes all of the high, medium, and low priority alarms and informational
messages. See the Troubleshooting section for more information.
Alarm
Priority
Loss of Power
High
Ventilator
High
Inoperative
Audible
Visual (Alarm
Indicator Button
and Display)
Red flashing button;
Blank screen
Red flashing
button; "Ventilator
Inoperative" message
Device
User/Clinician Action
Action
Shuts down
User: Press Start/Stop button
and then press Right button
to silence alarm. If using
AC power, try plugging
device into alternate AC
power source. If loss of
power continues, switch to
DC power by connecting
a fully charged detachable
or external battery to the
device. If there is still no
power, connect patient
to alternate source of
ventilation and contact your
health care professional.
Clinician: Restore AC power.
If alarm continues, connect a
fully charged detachable or
external battery to device to
restore power. If alarm still
continues, have the device
serviced.
User: Press Start/Stop button.
Shuts down if
If display is operational,
can't provide
Power Off confirmation
therapy safely.
screen appears. Select
Or, continues
Right button to shut off
to operate at a
device and silence alarm.
limited level.
Immediately remove patient
from ventilator and connect
them to alternate source of
ventilation. Contact your
health care professional for
service.
Clinician: Restore power. If
alarm continues, have the
device serviced.
Chapter 6 Ventilator Alarms