Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
IEC 60601 Test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
outside ISM bands
10Vrms
150 kHz to 80 MHz
in ISM bands
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the location in which the device is used exceeds the applicable
RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
Compliance
Level
Level
3 V
a
10 V
a
10 V/m
26 MHz to 2.5 GHz
Electromagnetic Environment -
Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d =
1.2
P
d =
1.2
P
d =
1.2
P
80 MHz to 800 MHz
d =
2.3
P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
the compliance level in each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Chapter 13 EMC Information
157
, should be less than