System Checkout
Remote Alarms
Trilogy100 clinical manual
Do not use the ventilator on a patient until a system checkout has been
performed. See Chapter 10 of this manual.
To make sure the device is operating properly at start-up, always verify
that the audible tone sounds and the alarm LEDs light red and then
yellow momentarily. Contact Philips Respironics or an authorized
service center for service if these indications do not occur at start-up.
When using a remote alarm, make sure you fully test the remote alarm
connector and cable by verifying that:
Annunciated alarms on the ventilator are also
annunciated on the remote alarm.
Disconnecting the remote alarm cable from the
ventilator or from the remote alarm results in an alarm
notification at the remote alarm.
The remote alarm should be tested daily.