PHILIPS Respironics Trilogy 100 Clinical Manual page 77

Table of Contents

Advertisement

70
Mode Setting
You can change the Mode setting on the Settings and Alarms screen
to one of the following therapy modes:
CPAP
S
S/T
T
PC
PC-SIMV
CV
AC
SIMV
Circuit Type
There are two circuit types you can select:
Passive
Active PAP
The Passive circuit uses the a passive exhalation device. The Active
PAP circuit uses an active exhalation device with a proximal air
pressure sensing connection.
When using the Passive circuit, the ventilator displays estimated
patient pressures based on the resistance of the standard patient
circuit (Whisper Swivel II with 1.8 meter tubing). Adding accessories
to the patient circuit (humidifier, water trap, etc.) may cause an
increase in circuit resistance and cause the device to display slightly
higher pressures than what is actually delivered to the patient.
With the Active PAP circuit, patient pressure is measured directly and
is not affected by any change in circuit resistance.
Trilogy100 clinical manual
Note: See Chapter 3 for detailed
information about each therapy
mode.
Note: Refer to the chart in
Chapter 3 to easily review all of
the settings available in each
therapy mode.
Note: This chapter describes
how to enable all of the device
settings, including device
alarms. For detailed information
about each alarm, please refer
to Chapter 6.
Note: To change the circuit
type, you must be in the
Setup screen with the airflow
turned off. See the Full Menu
Access Mode section for more
information.
Note: When the Circuit Type
setting is set to Passive Circuit,
all Ramp Start Pressure settings
in all modes will maintain the
minimum range.
Note: When the Circuit Type
setting is set to Active PAP
Circuit, the Flex and AVAPS
features are unavailable.