10. System Checkout Procedures
This chapter details the test procedures that should be performed by
the clinician prior to connecting the device to the patient. Test both
the active and passive circuits if you want to do a complete checkout
on the device. The tests should be performed as described in order
to verify proper operation of the device. Some of the procedures in
this chapter require you to change settings on the device. If you are
not familiar with the procedures for accomplishing this, please refer
to Chapter 4, Ventilator Setup and Chapter 5, Viewing and Changing
Settings.

Tools Required

Active Exhalation Porting Block with PAP
Passive Exhalation Porting Block
Active PAP Exhalation Device
Passive Exhalation Device
Test Lung
Small Flat Head Screwdriver

Visual Inspection

1. Verify that the enclosure is not broken and that all applicable
screws are in place.
2. Verify that the device handle, SD Card door, and detachable
battery are secure and in good working order.
Trilogy100
clinical manual
Note: The actual circuit
configuration to be used on
the patient should be used to
perform the system checkout
procedure.
WARNING
If you notice any unexplained
changes in the performance
of the device, if it is making
unusual sounds, if the device
or detachable battery are
dropped, if water is spilled into
the enclosure, or if the enclosure
is cracked or broken, discontinue
use and contact Philips
Respironics or an authorized
service center for service.
Chapter 10 System Checkout Procedures
133