Updating Prescriptions Using The Sd Card - PHILIPS Respironics Trilogy 100 Clinical Manual

Table of Contents

Advertisement

84

Updating Prescriptions Using the SD Card

With the Trilogy100, you can update the patient's prescription using
the SD Card. This feature lets you update a single prescription or
both prescriptions if the dual prescription feature is enabled. The
prescription update can occur either when the ventilator is off or on.
1. Insert an SD Card with a valid prescription into the device. A
"Change Prescription?" message appears on the display.
2. Select No to cancel the prescription update process and return
to the previous display (the black screen if the airflow was off or
the Monitor/Standby screen if the airflow was on). Select Yes to
start the prescription update process. Once the prescription is
read in and validated, one of the following screens appears on
the display to allow you to ensure the prescription is correct:
Trilogy100 clinical manual
WARNING
When you change the device
prescription, alarms and other
settings using the SD card,
Trilogy100 requires that the
caregiver review and verify the
changes prior to the changes
being used by the device.
The caregiver or health care
professional is responsible to
ensure that the prescription
settings are correct and
compatible with the patient
after using this feature. Installing
the wrong prescription for a
particular patient may result
in improper therapy, lack of
appropriate safety monitoring,
and risk of death or injury to the
patient.
Ventilator Off Prescription
Display Screen
Ventilator On Prescription
Display Screen

Advertisement