Emerson Rosemount 5408 Reference Manual page 185

Level transmitter, non-contacting radar with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 5408:
Table of Contents

Advertisement

Reference Manual
00809-0300-4408
Table B-9: For Divisions:
Temperature class /
Maximum surface
temperature
Division Gas groups:
T2
T3
T4
Division Dust groups:
T3
T4
T5
T6
Table B-10: For Zones:
Temperature class /
Maximum surface
temperature
Zone Gas groups:
T2
T3
T4
Zone Dust groups:
T250°C
T200°C
T135°C
T100°C
T85°C
B.9.3
IF FISCO
Certificate FM-C FM16CA0011X
Standards C22.2 NO. 0.4-17:2017, C22.2 NO. 0.5-16:2016, C22.2 No. 25-17:2017,
Markings
Rosemount 5408 Level Transmitter
Ambient temperature range
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
Ambient temperature range
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
-55 °C ≤ Ta ≤ 70 °C
C22.2 No.94-M91:1991 (R:2011), C22.2 No. 213-16:2016, C22.2 No.
61010-11:2004, CAN/CSA C22.2 No. 60079-0:2015 Ed. 3, CAN/CSAC22.2
No. 60079-11:2014 Ed. 2, CAN/CSAC22.2 No. 60079-15:2016 Ed.2, C22.2
No. 60079-26:2016, C22.2. 60529:2016; ANSI/ISA 12.27.01:2011
IS CL I, ll, lll DIV 1, GRPS A-G T4...T2
NI CL I, DIV 2, GRPS A-D T4...T2
Product Certifications
June 2019
Process temperature range
-55 °C to 250 °C
-55 °C to 195 °C
-55 °C to 130 °C
-55 °C to 160 °C
-55 °C to 130 °C
-55 °C to 95 °C
-55 °C to 80 °C
Process temperature range
-55 °C to 250 °C
-55 °C to 195 °C
-55 °C to 130 °C
-55 °C to 250 °C
-55 °C to 195 °C
-55 °C to 130 °C
-55 °C to 95 °C
-55 °C to 80 °C
185

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents