Reference Manual
00809-0300-4408
5.7.2
Run verify level using AMS Device Manager and
handheld communicator
Verify and adjust level measurement by comparing it to a hand-gauged measurement.
Prerequisites
Note
Before running Verify Level, make sure that: the product surface is calm, the tank is not
being filled or emptied, and the actual level is well above the tank bottom.
Procedure
1. Select Configure → Guided Setup.
2. Select Verify Level to check your level measurement, and follow the on-screen
5.8

Write protect a transmitter

5.8.1
Write protect a transmitter using Rosemount Radar
Master Plus
The transmitter can be write protected to prevent unauthorized changes.
Procedure
1. Under Overview, select Device Information → Security.
2. Select Change Write Protection and follow the on-screen instructions.
5.8.2
Write protect a transmitter using AMS Device Manager
and handheld communicator
The transmitter can be write protected to prevent unauthorized changes.
Procedure
1. Select Configure → Manual Setup → Device Setup → Security.
2. Select Change Write Protection and follow the on-screen instructions.
Rosemount 5408 Level Transmitter
instructions.