Emerson Rosemount 5408 Reference Manual page 116

Level transmitter, non-contacting radar with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 5408:
Table of Contents

Advertisement

Service and troubleshooting
June 2019
2. Select Restore Configuration.
3. Browse to the backup file and select Open.
7.5.5
Restore to default settings
Restore to default settings using Rosemount Radar Master
Plus
This function restores the transmitter to default settings (user configuration is
overwritten).
Prerequisites
Before restoring the transmitter to default settings, it is recommended to backup the
device configuration. The backup file can be used to restore configuration at a later stage.
Procedure
1. Under Service Tools, select Maintenance → Reset/Restore.
2. Select Restore Default Settings or Restore Fieldbus Default Settings, and follow
Related information
Save a backup file of the device configuration
Restore to default settings using AMS Device Manager and
handheld communicator
This function restores the transmitter to default settings (user configuration is
overwritten).
Prerequisites
Before restoring the transmitter to default settings, it is recommended to backup the
device configuration. The backup file can be used to restore configuration at a later stage.
Procedure
1. Select Service Tools → Maintenance → Reset/Restore.
2. Select Restore Default Settings or Restore Fieldbus Default Settings, and follow
116
the on-screen instructions.
Option
Restore Default Settings
Restore Fieldbus Default Settings
the on-screen instructions.
Description
Resets the configuration to default settings
(user configuration is overwritten). The
F
Fieldbus settings are preserved.
OUNDATION
Resets the F
Fieldbus settings
OUNDATION
(Resource, Transducer, and Function
Blocks) to default settings.
Reference Manual
00809-0300-4408
Reference Manual

Hide quick links: