Emerson Rosemount 5408 Reference Manual page 115

Level transmitter, non-contacting radar with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 5408:
Table of Contents

Advertisement

Reference Manual
00809-0300-4408
Suppress false echoes using AMS Device Manager
Procedure
1. Select Service Tools → Echo Tuning → Suppress. Suppressed echoes are shown in
2. Select Suppress or Remove Suppression.
3. Type the distance to the echo that should be added to or removed from the list, and
4. Select Finish when the Method Complete message appears.
Suppress false echoes using handheld communicator
Procedure
1. Select Service Tools → Echo Tuning → Suppress.
2. To view the current suppressed echoes, select Suppressed Echoes.
3. Select Suppress or Remove Suppression.
4. Type the distance to the echo that should be added to or removed from the list, and
7.5.3
Save a backup file of the device configuration
When configuration is finished, it is recommended to store the device configuration in a
backup file for future reference using Rosemount Radar Master Plus. A backup of the
device configuration will be saved to file as well as a configuration report (optional).
The backup file may be useful to:
Restore the configuration of the transmitter.
Install another transmitter in a similar tank.
Troubleshoot the transmitter.
Procedure
1. In Rosemount Radar Master Plus, under Service Tools, select Maintenance →
2. Select Save Configuration.
3. Type your desired file name.
4. Browse to the desired directory, and then select Save.
5. Optional: Select the Create and save report (.pdf) check box.
6. Select Save.
7.5.4
Download configuration from file to device
Procedure
1. In Rosemount Radar Master Plus, under Service Tools, select Maintenance →
Rosemount 5408 Level Transmitter
the table.
then select Next.
then select Enter.
Backup.
Backup .
Service and troubleshooting
June 2019
115