Additional Certifications - Emerson Rosemount 5408 Reference Manual

Level transmitter, non-contacting radar with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 5408:
Table of Contents

Advertisement

Product Certifications
June 2019
2. Plastic wire-on tag, Plastic part of Process Seal Antenna and Non-standard paint
3. Cable entries must be used which maintain the ingress protection of the enclosure
4. The applicable temperature class, ambient temperature range and process
Temperature class /
Maximum surface
temperature
T2
T3
T4
B.12

Additional certifications

B.12.1
Suitable for intended use
Compliant with NAMUR NE 95:2013, "Basic Principles of Homologation".
B.12.2
QA 3-A
Certificate Authorization Number
The following options are conforming to the 3-A Sanitary Standards, Number 74-06
(Sensors and Sensor Fittings and Connections):
Process connection type
Process connection size
Antenna type
Antenna size
The certification of the transmitter relies upon the following materials used in its
construction:
Table B-13: Product Contact Surfaces
Item
Microwave launcher
196
options (paint options other than Rosemount Blue) may cause risk from
Electrostatic discharge. Avoid installation that could cause electrostatic build-up,
and only clean with a damp cloth.
to at least IP65. To maintain the ingress protection ratings, Covers and Sensor
Module to be fully tightened and PTFE tape or pipe dope is required for cable entries
and blanking plugs. See Instruction Manual on application requirements.
temperature range of the equipment is as follows:
Ambient temperature range
-34 °C ≤ Ta ≤ 70 °C
-34 °C ≤ Ta ≤ 70 °C
-34 °C ≤ Ta ≤ 70 °C
®
C (Tri-Clamp
2, 3, 4
SAA (Process Seal antenna)
2, 3, 4
Material
PTFE fluoropolymer
Process temperature range
-34 °C to 250 °C
-34 °C to 195 °C
-34 °C to 130 °C
3626
®
)
Reference Manual
00809-0300-4408
Reference Manual

Hide quick links:

Advertisement

Table of Contents
loading