Replacing The Operating Panel Locking Wire; Replacing The Up/Down Slide Wire - Toshiba Aplio SSA-770A Service Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
3.24
Replacing the Operating Panel Locking Wire
3.24.1
Replacing the up/down slide wire
Left
* Only for the SSA-770A
(1)
Removing the covers
(a)
Remove the two caps <A>.
(b)
Remove the two screws CA> (M4) and remove the right side cover.
(c)
Remove the two screws cB> (M4) and remove the right side cover for the
peripheral units.
(d)
Remove the two caps cB=-.
(e)
Remove the two screws CC> (M4) and remove the left side cover.
(f)
Remove the two screws CD> (M4) and remove the left side cover for the peripheral
units.
(g)
Remove the four screws cE> (M4) and remove the tray.
(h)
Remove the four screws cF> (M4) and remove the top front cover.
Right side cover for
the peripheral units
Right side cover
Screw cB>
I
Left side cover for
the peripheral units
Q
T
Cap <A>
Screw CA>
(SBPPC12SCR)
(SBPP4-12SCR)
Screw cE>
(SBPP4-12SCR)
I
'IS
crew CC>
(SBPP4-12SCR)
Cap'cB>
3-67