No. 2D730-148E*O
3.2.2
For the SSA-700A
(1)
Removing the left side cover
(a)
Remove the two rubber caps at the bottom of the cover.
(b)
Remove the two M4 screws <A> to remove the left side cover.
* The cover can be removed by pulling the bottom of the cover toward you
(2)
Removing the right side cover
Remove the right side cover in the same manner as the left side cover.
(3)
Removing the front cover
When the transducers
are connected.
remove them.
It is necessary to remove the right and left side covers in advance.
(a)
When the blank cover is installed, loosen the M3 screw cB> to remove the blank
cover with the printer base installed.
(b)
Remove the four M4 screws CC> to remove the front cover
(4)
Removing the rear cover for the power supply
It is necessary to remove the right and left side covers in advance.
Remove the four M4 screws CD> to remove the rear cover for the power supply.
(5)
Removing the base front cover
It is necessary to remove the right and left side covers and the front cover in advance.
(a)
Remove the two M4 screws cE>.
(b)
Loosen the two M4 screws cF> to remove the base front cover.
(6)
Removing the base rear cover
It is necessary to remove the right and left side covers and the rear cover for the power
supply in advance.
Loosen the two M4 screws cG> to remove the base rear cover.
* The cover can be removed by pulling it toward you slightly and lifting it obliquely.
(7)
Removing the peripheral unit rack
Remove the six M4 screws <H> to remove the rack.