Toshiba Aplio SSA-770A Service Manual page 183

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
(i)
Close the service mode menu.
Press [-] or [+I.
Press [v].
The service mode menu closes.
Click the [Close] button on the Adjust the Range Automatically
screen shown in (h).
(j)
Select [Start] + [Shutdown...]
+ [Shutdown] to shut down the system.
(k)
Restart the system and confirm that there are no abnormalities
in LCD monitor display.
CAUTION:
This image quality adjustment
procedure is applicable only to the
U I LM-770A/700A.
Auto Adjustment
and Range adjustment
must be performed without fail.
7-16
*