Imaging Preset; 3D Preset - Toshiba Aplio SSA-770A Service Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
(5)
Imaging preset
The imaging preset includes a group of parameters
related to imaging.
They are available to
the user. However, there is no preset editor.
User setting can be performed using the Save
function after selecting the imaging parameter.
Up to seven imaging presets can be
registered for each Exam Type by the user and can be deleted and overwritten.
Note,
however, that the factory setting that is contained in each Exam Type cannot be deleted or
overwritten.
Imaging parameters
can be preset independently
for Fund., THI, CHI, Color, Power, etc.
The 2D-IP is equivalent to B-IP in the previous system. There are five IP Types (A to E) and
eight IP Indexes (1 to 8). Parameters such as the dynamic range value in 2D-IP are stored
separately (in a separate file) from the imaging preset parameters.
(6)
3D preset
The 3D Preset consists of the following three presets. Although these presets have been set
at the factory, they can be changed by the user. These presets can be set for each Exam
Type and each Volume Type.
l
Initial display preset
0 Operation preset
l
Reconstruction
preset
11-2